Filtered by vendor Hospira Subscriptions
Filtered by product Lifecare Pca5 Subscriptions
Total 6 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2015-3459 1 Hospira 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware 2017-01-03 N/A
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
CVE-2015-3958 1 Hospira 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware 2016-12-06 N/A
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.
CVE-2015-3957 1 Hospira 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware 2016-12-06 N/A
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
CVE-2015-3955 1 Hospira 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware 2016-12-06 N/A
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
CVE-2015-1011 1 Hospira 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware 2015-07-08 N/A
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.
CVE-2014-5406 1 Hospira 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware 2015-07-08 N/A
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.