The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
References
Link | Resource |
---|---|
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm | Third Party Advisory US Government Resource |
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 | Third Party Advisory US Government Resource |
https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/ |
History
No history.
MITRE Information
Status: PUBLISHED
Assigner: icscert
Published: 2015-07-06T19:10:00
Updated: 2015-07-06T18:57:01
Reserved: 2014-08-22T00:00:00
Link: CVE-2014-5406
JSON object: View
NVD Information
Status : Analyzed
Published: 2015-07-06T19:59:00.097
Modified: 2015-07-08T15:18:06.177
Link: CVE-2014-5406
JSON object: View
Redhat Information
No data.
CWE