A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.
History

No history.

cve-icon MITRE Information

Status: PUBLISHED

Assigner: mitre

Published: 2021-08-02T12:43:57

Updated: 2021-08-02T12:43:56

Reserved: 2021-07-21T00:00:00


Link: CVE-2021-37160

JSON object: View

cve-icon NVD Information

Status : Modified

Published: 2021-08-02T13:15:07.707

Modified: 2023-11-07T03:36:55.140


Link: CVE-2021-37160

JSON object: View

cve-icon Redhat Information

No data.

CWE