Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
History

No history.

cve-icon MITRE Information

Status: PUBLISHED

Assigner: icscert

Published: 2020-06-29T13:49:50

Updated: 2020-06-29T13:49:50

Reserved: 2020-04-21T00:00:00


Link: CVE-2020-12035

JSON object: View

cve-icon NVD Information

Status : Analyzed

Published: 2020-06-29T14:15:11.397

Modified: 2020-07-14T15:34:03.733


Link: CVE-2020-12035

JSON object: View

cve-icon Redhat Information

No data.