A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
References
Link | Resource |
---|---|
http://www.securityfocus.com/bid/100665 | Third Party Advisory VDB Entry |
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A | Third Party Advisory US Government Resource |
History
No history.
MITRE Information
Status: PUBLISHED
Assigner: icscert
Published: 2018-02-15T10:00:00
Updated: 2018-02-15T10:57:01
Reserved: 2017-08-09T00:00:00
Link: CVE-2017-12725
JSON object: View
NVD Information
Status : Analyzed
Published: 2018-02-15T10:29:00.523
Modified: 2018-03-02T14:40:15.310
Link: CVE-2017-12725
JSON object: View
Redhat Information
No data.
CWE