A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
References
Link Resource
http://www.securityfocus.com/bid/100665 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A Third Party Advisory US Government Resource
History

No history.

cve-icon MITRE Information

Status: PUBLISHED

Assigner: icscert

Published: 2018-02-15T10:00:00

Updated: 2018-02-15T10:57:01

Reserved: 2017-08-09T00:00:00


Link: CVE-2017-12725

JSON object: View

cve-icon NVD Information

Status : Analyzed

Published: 2018-02-15T10:29:00.523

Modified: 2018-03-02T14:40:15.310


Link: CVE-2017-12725

JSON object: View

cve-icon Redhat Information

No data.

CWE