Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
Link | Resource |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 | Mitigation US Government Resource Third Party Advisory |
History
No history.
MITRE Information
Status: PUBLISHED
Assigner: icscert
Published: 2019-03-25T17:44:44
Updated: 2019-03-25T17:44:44
Reserved: 2015-05-12T00:00:00
Link: CVE-2015-3956
JSON object: View
NVD Information
Status : Modified
Published: 2019-03-25T18:29:00.323
Modified: 2019-10-09T23:14:05.973
Link: CVE-2015-3956
JSON object: View
Redhat Information
No data.
CWE