Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
Link | Resource |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation US Government Resource Third Party Advisory |
History
No history.
MITRE Information
Status: PUBLISHED
Assigner: icscert
Published: 2019-03-26T15:17:16
Updated: 2019-03-26T15:17:16
Reserved: 2014-08-22T00:00:00
Link: CVE-2014-5432
JSON object: View
NVD Information
Status : Modified
Published: 2019-03-26T16:29:00.290
Modified: 2019-10-09T23:11:10.730
Link: CVE-2014-5432
JSON object: View
Redhat Information
No data.