Filtered by CWE-1299
Total 4 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2023-29063 2 Bd, Hp 3 Facschorus, Hp Z2 Tower G5, Hp Z2 Tower G9 2023-12-05 2.4 Low
The FACSChorus workstation does not prevent physical access to its PCI express (PCIe) slots, which could allow a threat actor to insert a PCI card designed for memory capture. A threat actor can then isolate sensitive information such as a BitLocker encryption key from a dump of the workstation RAM during startup.
CVE-2023-29060 2 Bd, Hp 3 Facschorus, Hp Z2 Tower G5, Hp Z2 Tower G9 2023-12-05 5.7 Medium
The FACSChorus workstation operating system does not restrict what devices can interact with its USB ports. If exploited, a threat actor with physical access to the workstation could gain access to system information and potentially exfiltrate data.
CVE-2022-43557 1 Bd 14 Bodyguard 121 Twins, Bodyguard 121 Twins Firmware, Bodyguard 323 Colorvision and 11 more 2023-10-27 5.3 Medium
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
CVE-2021-3788 1 Binatoneglobal 42 Cn28, Cn28 Firmware, Cn40 and 39 more 2022-10-27 6.8 Medium
An exposed debug interface was reported in some Motorola-branded Binatone Hubble Cameras that could allow an attacker with physical access unauthorized access to the device.