Filtered by vendor Medtronic Subscriptions
Filtered by product 24952 Mycarelink Monitor Subscriptions
Total 2 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2018-8870 1 Medtronic 4 24950 Mycarelink Monitor, 24950 Mycarelink Monitor Firmware, 24952 Mycarelink Monitor and 1 more 2019-10-09 N/A
Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
CVE-2018-8868 1 Medtronic 4 24950 Mycarelink Monitor, 24950 Mycarelink Monitor Firmware, 24952 Mycarelink Monitor and 1 more 2019-10-09 N/A
Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.